Measuring principle Headspace gas chromatography
Measuring range 0 - 999 ppm* methanol
Accuracy** 5 ppm or 5% of the measured value (whatever is larger)
Resolution 1 ppm
Limit of detection 1 ppm
Ethanol limit 0 - 20 vol% ethanol in sample
*The device informs the user when the concentration exceeds 999 ppm. Mass concentratrion (DIN 1310) at a reference temperature of 25 °C.
**Standard deviation of the reproducibility with methanol-water mixtures (10 - 1000 ppm)